The UDT TRAMP is a transimpedance amplifier (current-to-voltage) instrument that provides a low input impedance to accurately measure the short circuit current of photo transducers such as silicon and germanium photodetectors, vacuum photodiodes and photomultiplier tubes.
The TRAMP has been specifically designed as a low cost, lab quality instrument to interface silicon photodiodes to a variety of measurement equipment. This model provides multiple gain selection and utilizes a common BNC connector for input and output connections for user convenience. A voltmeter, oscilloscope, chart recorder or any other voltage sensitive instrument may be used to monitor the amplifier output.
Key Applications
- Laboratory Research
- Product Line Integration
- Automatic Test Equipment
- Laser Pulse Measurement
- Eight decades of gain ranging between 10³ and 1010
- Remote computer control using data acquisition interface
- Integral NiCad battery pack with normal & trickle charge modes
- Line powered operation using supplied transformer
- Output scaling to interface to chart recorders and other equipment
- Optimal measurement range indicator
UDT Tramp General Specifications | |
---|---|
Gain (Volt / Ampere) | 103 to 1010 |
Noise (rms) | 0.5 mV |
Current Range | 10 -2 to 10 -13 Amps |
Overall Accuracy | ± 2% |
Bandwidth vs Gain | 103 @ 160 kHz 104 @ 45 kHz 105 @ 12 kHz 106 @ 12 kHz 107 @ 550 Hz 108 @ 550 Hz 109 @ 5 Hz 1010 @ 5 Hz |
Offset Drift vs Temperature | < 50 µV / °C |
Input Impedance | < 1 Ω |
Output Voltage | ± 5 V |
Input Power | 115 / 230 VAC |
Battery Information | 10 Each, AA NiCad Typical operating time of 15 hours |
Dimensions | 70 mm (2.8 in) H x 150 mm (5.9 in) W x 191 mm (7.5 in) D |
Calibration Traceability | Traceable to NIST with optional ISO/IEC 17025 accredited |
Relative Humidity | Up to 99% (non-condensing) |
Regulatory Compliance | TUV, UL, CSA, CE |
Specifications are subject to change without notice. |